Starting at $9,540.00
  • Usually ships in 3-4 weeks

Overview

Features

  • 0.10ppb to 10ppm
  • ±5% of reading or ±0.1ppb to (whichever is greater) within ±5°C of calibration temperature accuracy
  • ± 5% of reading or ± 0.1ppb (whichever is greater) at constant temperature reproducibility
  • 90% of 10 – 1ppb step: less than 3 minutes; 90% of 100 – 1ppb step: 5 minutes response time
  • Automatic temperature compensation
  • Alkaline vapor buffer:0.88 SG (28% to 30% NH3) ammonia above 0.5ppb sodium, 99% diisopropylamine solution below 0.5ppb sodium pH compensation
  • Automatic 2-point or single point; user may select weekly, bi-monthly, or monthly calibration frequency as well as day and time calibration
  • Two isolated 4-20mA current (analog) outputs
  • Large user friendly color touch screen interface
  • Multi-stream available
  • Calibration: automatic 2-point or single point; user may select weekly, bi-monthly, or monthly calibration frequency as well as day and time
  • Hassle-free maintenance: calibration standards and electrode refill can last up to 3 months
  • Low sample flow indication (optional)
  • Quick connect electrodes and probe positioning clamps
  • Automatic reference electrode electrolyte refill system (standard in US and optional elsewhere)
  • Grab sample capability
  • Internal data logger set to store readings every 10 minutes for 30 days
  • External data logger can be set for every 1 minute (optional)
  • Automatically logs last 10 alarms and results of last 10 calibrations
  • Reagents: Ammonium Hydroxide (standard) or Di-isopropylamine (optional)

Description

The Waltron 9033X Sodium Analyzer delivers exceptional results in a self-contained package. An enhanced color touchscreen user interface provides fully-automatic calibration, ease of operation and maintenance, and increased functionality. The 9033X can be used in various ultrapure water applications, providing reliable and accurate control of sodium independent of flow. Common measuring points in a boiler system for online sodium measurement include make-up water, condensate, boiler feedwater, and saturated steam/main steam.

Applications:

  • Boiler water
  • Cooling water