Panametrics MISP2 Probe
MISP2 aluminum oxide probe for use with Panametrics moisture.IQ and PM880 moisture analyzers.
Overview
Features
- Intrinsically safe
- Ambient to ppb moisture measurement
- 16-bit resolution
- Optional built-in temperature and pressure sensors
- Nonvolatile calibration data storage
- Calibrations traceable to NIST
- Requires only twisted pair cabling
- Threaded and VCR mount configurations
If you are replacing an existing MISP probe, MISP2 adapter cable is required
This product is an accessory for the following products:
Description
The Panametrics aluminum oxide MISP2 probe has set the standard of performance and value in industrial moisture measurement for more than 40 years.
In use, the MISP2 Probe is coupled to Panametrics hygrometer consoles by an interconnecting cable. Ease of use, wide measurement range and rigorous calibration standards make these systems the preferred choice for industrial moisture measurement worldwide.
Available Built-In Pressure and Temperature Measurement
Accurate determination of many moisture measurement parameters requires knowledge of the process temperature and pressure. The inconvenience and limitations associated with installing and using separate temperature and pressure sensors have been eliminated, with both capabilities being built directly into the Moisture Image Series probe. A nonlinear NTC thermistor to measure temperature from -30 to 70°C (-22 to 158°F) and a choice of five solid-state piezoresistive transducers, to measure pressures up to 5,000 psig (345 bar), are available.
Temperature and pressure input data are used by moisture.IQ analyzer and the PM880 portable hygrometer to determine parameters such as ppm, lbs/mmSCF and releative humidity.
Available Rigorous Calibration Traceable to the NIST
The aluminum oxide sensor of each MISP2 probe is individually calibrated in one fo the world's most advanced moisture calibration facilities. Developed over several decades, this facility generates precisely known moisture concentrations, traceable to the National Institure of Standards and Technology (NIST), to which each sensor is exposed during the calibration process.
All data is gathered and stored by a dedicated computer system. Calibrations are repeated over a period of many months to ensure the stability of each individual moisture probe. Only those probes which meet Panametrics Infrastructure Sensing's demanding specifications for accuracy and stability are shipped to customers.
Installation Flexibility
The MISP2 probe is designed to be located at the process, exactly where the measurement is needed. The sensor can withstand process temperatures of -110 to 70°C (166 to 158°F) and pressures from vacuum up to 5,000 psig (345 bar). No minimum flow rate is required. For additional flexibility, the probe can be located up to 3,000 feet (0.9 km) from the analyzer, connected by inexpensive, unshielded, twisted-pair cable (AWG 22). As a result, the probe can be placed in the ideal location, even in a large plant, without the cost, delays and installation problems associated with special cabling.