The MadgeTech ULT90 Ultra Low Temperature Freezer Validation System can measure and log data as
low as -86°C. It can be placed in small and large freezers of varying configurations.
The system provides a complete stand-alone temperature mapping and validation
solution for freezers to ensure uniform storage and testing conditions. The ULT90
includes a USB docking station with 6 feet of cable (IFC300), the Cryo-Temp data
logger, and MadgeTech's FDA 21 CFR Part 11 Secure Software and IQ/OQ/PQ validation
protocols. For larger freezers, loggers can be plotted in one graph to show
variances in different parts of the freezer. Finally, as failure to monitor
external freezer temperatures may invalidate internal freezer temperatures,
an additional Cryo-Temp can be included for placement outside of the freezer.
Please note, the ULT90-9 is not compatible with 64 bit operating systems.
Applications for the MadgeTech ULT90:
- Ultralow Freezer Validation
- Ultralow Freezer Mapping
- Ultralow Freezer Stabilization Studies
- Temperature Surveys
- Reveal stratification of air
- Reveal hot spots
- Reveal anomalies in chilling systems