Martel MC-1010 Multi-Function Calibrator
High accuracy, multi-function process calibrator.
- 1919999
Overview
Features
- Simple, straight forward, Intuitive user interface
- Easy to learn yet powerful 3 Key Martel Menu System
- New ClearBrite Graphic Display
- New streamlined case design
- ±0.4°C thermocouple/±0.3°C RTD accuracy
- ±0.015% of reading electrical accuracy
- Simulates 13 T/C types and 13 RTD types, plus YSI 400
- Read and source modes
- Full Function 4 to 20 mA read, source, simulate and read/power modes with ±0.015% of reading accuracy
- Source or read 0 to 20 volts (±0.015% of reading accuracy)
- Direct keyboard entry enables quick output changes
- Store up to nine (9) setpoints for each output function
- RTD function compatible with all pulsed transmitters
- RS232 serial interface
- Includes NIST Certificate with calibration data
- Communication port compatible with both Fluke 700 Series and Beta Calibrator pressure modules
Description
The Martel MC-1010 is a multi-function process calibrator that combines a high accuracy process calibrator with common DMM functions. With the ability to accurately calibrate virtually any process, the Martel MC-1010 process calibrator is an indispensible tool for a wide range of applications.
The Martel MC-1010 multi-function process calibrator will test continuity as well as read and source thermocouples, RTDs, Ohms, voltage, current, and frequency. The unit is compatible with BetaPort-P and Fluke 700 series pressure modules which expand its range to pressure applications. Results are shown of the high contrast ClearBrite graphic display which features a vivid white backlight that makes the display easy to read in all light conditions. Martel's easy to learn yet powerful 3 key menu structure provides a highly intuitive yet powerful operator interface.
The Martel MC-1010 multi-function process calibrator can store up to 9 setpoints for each output function. An RS232 serial interface provides communications options.
The Martel MC-1010 multi-function calibrator ships complete with:
- (4) AA Alkaline batteries
- Full Rubber boot
- (2) Test leads
- Instruction manual
- NIST Certificate with calibration data
DC Voltage
Display | Range | Accuracy |
---|---|---|
Non-isolated | 0.000 to 20.000 V | 0.015% ± 2 mV |
Source | 0.000 to 20.000 V | 0.015% ± 2 mV |
DC Current
Display | Range | Accuracy |
---|---|---|
Non-isolated | 0.000 to 24.000 mA | 0.015% ± 2 µA |
Source | 0.000 to 24.000 mA | 0.015% ± 2 µA |
Frequency
Display | Range | Accuracy |
---|---|---|
Measurement | 2.0 to 600.0 CPM | 0.05% ± 0.1 CPM |
1.0 to 1000.0 Hz | 0.05% ± 0.1 Hz | |
1.00 to 10.00 kHz | 0.05% ± 0.01 kHz | |
Source | 2.0 to 600.0 CPM | 0.05% |
1.0 to 1000.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.125% | |
1.00 to 10.00 kHz | 0.125% |
Resistance
Display | Range | Excitation Current | Accuracy |
---|---|---|---|
Measurement | 0.00 to 400.0 Ω | 0.025% ± 0.05 Ω | |
401.0 to 4000.0 Ω | 0.025% ± 0.5 Ω | ||
Source | 5.0 to 400.0 Ω | 0.1 to 0.5 mA | 0.025% ± 0.1 Ω |
0.5 to 3 mA | 0.025% ± 0.05 Ω | ||
401 to 1500 Ω | 0.05 to 0.8 mA | 0.025% ± 0.5 Ω | |
1500 to 4000 Ω | 0.05 to 0.4 mA | 0.025% ± 0.5 Ω |
MilliVolts
Display | Range | Accuracy |
---|---|---|
Read | -10.000 to 75.000 mV | 0.02% ± 10 µV |
Source | -10.000 to 75.000 mV | 0.02% ± 10 µV |
Temperature
Device | Types (see brochure for ranges and accuracies) |
---|---|
RTD measurement/source | PT385, 10 Ω; PT385, 50 Ω; PT385, 100 Ω; PT3926, 100 Ω; PT3916, 100 Ω; PT385, 200 Ω; PT385, 500 Ω; PT385, 1000 Ω; NI120; Cu10; Cu50; Cu100; YSI400 |
mV/thermocouples | J, K, T, E, R, S, B, C, XK, BP, L, U, N |
Pressure
Range | Accuracy |
---|---|
Pressure measurement via BetaPort-P pressure modules (requires BPPA-100 Pressure Module Adapter). See brochure for ranges and accuracies |
General Specifications
Power Supply | 4 x AA batteries, alkaline or optional rechargeable |
Display | High contrast 128 x 64 pixel addressable graphic LCD w/daylight backlight |
Serial Communications | ASCII, RS-232 (requires optional Martel 1919069 serial cable or 1919896 USB cable) |
CE-EMC | EN50082-1 and EN55022: 1994 Class B |
Safety | CSA C22.2 No. 1010-1: 1992 |
Protection | IP54; Protected against misconnection to 250 VAC/VDC |
Operating Temperature | -10 to 50°C |
Dimensions | 8.5 x 4.8 x 2.1" |
Weight | 1.8 lbs |
Documents
Accessories
Cases
Accessories
Please consider these optional accessories.
Need Help? Call a Calibration engineer at 1-800-884-4967
We're open 8:00 am to 5:30 pm ET