Martel PSC-4010 Process Calibrator
High accuracy, portable process signal calibrator.
- 1919895
Overview
Features
- Simple, straight forward, Intuitive user interface
- Easy to learn yet powerful 3 Key Martel Menu System
- New ClearBrite™ Graphic Display
- New streamlined case design
- Reads and simulates process control signals voltage, current, and frequency
- Multiple input/output ranges for current, frequency, and voltage modes
- Built-in 24 VDC source for powering two wire transmitters
- Accuracy ± 0.015% of reading
- Includes NIST Certificate with calibration data
Description
The Martel PSC-4010 process calibrator is a hand-held, full function voltage and current calibrator with the capability to both read and simulate common process control signals.
The Martel PSC-4010 process calibrator incorporates multiple input/output ranges for both current and voltage operating modes. When operated in the current mode, the Martel PSC-4010 process calibrator can source, simulate, or read over a 0 to 24 mA range with 0.01 mA resolution. The Martel PSC-4010 process calibrator can also supply 24 VDC to power a two-wire transmitter while simultaneously reading back its output both in milliAmps or as a % of full scale. When operated in the voltage range, the Martel PSC-4010 process calibrator can read or source over two ranges, 0 to 20.00 V or 0 to 200.0 mV.
For your convenience, you can store up to three (3) calibration points in the memory of your Martel PSC-4010 process calibrator for each voltage or current range. These calibration points are non-volatile and will remain in memory until the user changes them.
The Martel PSC-4010 process calibrator ships complete with:
- (4) AA Alkaline batteries
- (1) Test lead
- Full Rubber boot
- Instruction manual
- NIST Certificate with calibration data
DC Voltage
Display | Range | Accuracy |
---|---|---|
Non-isolated | 0.000 to 20.000 V | 0.015% ± 2 mV |
Source | 0.000 to 20.000 V | 0.015% ± 2 mV |
DC Current
Display | Range | Accuracy |
---|---|---|
Non-isolated | 0.000 to 24.000 mA | 0.015% ± 2 µA |
Source | 0.000 to 24.000 mA | 0.015% ± 2 µA |
Frequency
Display | Range | Accuracy |
---|---|---|
Measurement | 2.0 to 600.0 CPM | 0.05% ± 0.1 CPM |
1.0 to 1000.0 Hz | 0.05% ± 0.1 Hz | |
1.00 to 10.00 kHz | 0.05% ± 0.01 kHz | |
Source | 2.0 to 600.0 CPM | 0.05% |
1.0 to 1000.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.125% | |
1.00 to 10.00 kHz | 0.125% |
MilliVolts
Display | Range | Accuracy |
---|---|---|
Read | 0000 to 90.000 mV | 0.02% ± 10 µV |
Source | 0.000 to 100.000 mV | 0.02% ± 10 µV |
General Specifications
Power Supply | 4 x AA batteries, alkaline or optional rechargeable |
Display | High contrast 128 x 64 pixel addressable graphic LCD w/daylight backlight |
Serial Communications | ASCII, RS-232 (requires optional Martel 1919069 serial cable or 1919896 USB cable) |
CE-EMC | EN50082-1 and EN55022: 1994 Class B |
Safety | CSA C22.2 No. 1010-1: 1992 |
Protection | IP54; Protected against misconnection to 250 VAC/VDC |
Operating Temperature | -10 to 50°C |
Dimensions | 8.5 x 4.8 x 2.1" |
Weight | 1.8 lbs |
Documents
Need Help? Call a Calibration engineer at 1-800-884-4967